Everything you need for automated redox titrations, including SO2 determination in wine, reducing sugars, peroxide values, vitamin C content, and much more.

  • EasyPlus Easy OX has everything you need for inflection-point Redox titrations
  • This versatile titrator has a platinum ring redox-sensing electrode making it the ideal instrument to meet the needs of food, beverage, galvanic and even environmental analyses

Specifications

Burette Functions

EasyPlus burette 20 mL (included)

Accuracy (mV)

0.1

Accuracy (Temperature)

0.2°C

Backlight Option

Yes

Blank Options

Variable or fixed value

Burette Sizes

10 mL, 20 mL (Standard Delivery), or 50 mL

Calibration Points

2

Certifications/Compliance

CE, CSA

Connector Types

BNC, USB A, USB B, Mini-DIN, RCA

Cycles Per Titration Run

Maximum number of samples in a series is 5. When choosing this option the mean value and RSD is calculated.

Data Management Details

Export via USB, USB Printer

Description

Easy Ox

Display Type

4.3 inch, 480 x 272 pixel RGB

Type

EasyPlus Automated Titrator

Equivalence Points

Detection of up to 3

For Use With (Application)

Potentiometric

Includes

EM40-BNC Sensor Redox mV, EasyPlus burette 20 mL, EasyPlus tubing set instrument, EasyStir GT, EasyPlus titration head GT, EasyPlus drying tube, valve, bottle heads, insert set titration head GT, burette cover set

Inputs

BNC, USB A, USB B, Mini-DIN, RCA

Keypad

Touchscreen

Languages

15

Material

PP GF30

Methods

Reduction and oxidation (redox) titrations

Model

Easy Ox

Power Supply

24 VDC / 1.25 A

Precision

0.5% RSD

Probe Type

Platinum ring (redox titrations)

Range (mV)

±2000 mV

Relative Humidity Range

Max. 80% at 31°C

Resolution

±0.1mV

Resolution (Temperature)

0.1°C

Sample ID

None, Fixed, Variable

Sample Size

User Defined

Stirrer Probe Speeds

Low, Medium, High, Maximum

Temperature Range (English) Ambient

41 to 104°F

Temperature Range (Metric)

0 to 100°C

Temperature Range (Metric) Ambient

5 to 40°C

Temperature Selection

°C / °F

Time and Date

Yes

Titration Process Control

Normal, Fast, Cautious, Very cautious, User defined

Titration Setup Wizard

iTitrate

Titration Types

Direct, back compensated, blank compensated

Updateable Firmware

Yes

Volume (Metric) Sample

100 mL Beaker

Warranty

12 months

Dimensions (L x W x H)

220 x 170 x 350 mm

Width (English)

6.7 in.

Width (Metric)

170 mm

Depth (English)

8.66 in.

Depth (Metric)

220 mm

Electrical Requirements

100 to 240 V ±10% 50/60Hz

Voltage

100 to 240 V ±10%

Frequency

50/60 Hz

Product Title