Learn More
Vernier Vernier Chloride Ion-Selective Electrode (ISE)

Shop All Vernier Software Products


Includes
Chloride Ion-Selective Electrode, 30mL bottle of Chloride Low Standard (10mg/L), 30mL bottle of Chloride High Standard (1000mg/L), Short-term ISE soaking bottle, User manual
- Combination-style, non-refillable, gel-filled electrode
- Solid-state membrane that does not need to be replaced
- Gives quick measurement of salinity of water samples
- Measures chloride levels in ocean saltwater or salt in food samples
- Interfering ions: CN-, Br-, I-, S2-
- Compatible with LabQuest™ 2, Original LabQuest, LabQuest Mini, Go!Link™, SensorDAQ™, LabPro™, EasyLink™, TI-Nspire Lab Cradle™ interfaces
- Electrode slope (log voltage vs. concentration): -56mV/decade
- Electrode resistance: 1 to 4 MOhm
- Minimum immersion: 1.1 in.
Ion-Selective Electrodes require good chemical technique and careful calibration to obtain accurate results; they are not recommended for middle school or elementary students.
Specifications
Temperature (Metric) Operating
0°C to 50°C
Diameter (Metric) Body
12 mm
Diameter (Metric) Cap
16 mm
Length (English) Body
6.1 in.
Length (English) Cable
39.3 in.
Length (Metric) Body
155 mm
Length (Metric) Cable
100 cm
pH Range
2 to 12
Concentration Measurement Range
10 to 20,000mg/L