Learn More
HORIBA ABX SAS Yumizen C1200 Clinical Chemistry Analyzer with Autoloader
Next generation system designed to provide high efficiency and accuracy with optimal workflow capabilities for laboratories processing up to 2 million plus tests annually.
Supplier: HORIBA ABX SAS 5300002212
Alert:
This item is a non-standard voltage; please ensure you are ordering the correct product. Please contact customer service for assistance: 1-800-766-7000.
Description
General Description:
- Analysis method types: Spectrophotometry, colorimetric and immunoassay Flexible
- Indirect ISE Na+, K+, and Cl- with reference electrode – reliable
- Throughput 1200 tests per hour Efficient workflow for labor savings
- Automatic maintenance during start-up and shutdown for labor savings
Sample Handling:
- Sample tray holds 84 samples, continuous loading
- Sample probes liquid level sensing, cap/shock and clot detection
- Sample ID with Integrated barcode reader.
- Sample types: Serum, plasma and urine
- Sample collection format: 5, 7, 10 mL tubes and sample cups
- ISE sample volume: 1-25 uL (0.1 uL increment)
- Automatic sample dilution 1/2 to 1/200
- Calibration and Control tray: 61 refrigerated positions for calibrators and controls
Reagent Handling:
- Reagent tray: 2 trays with 45 reagent positions each
- Refrigeration: all reagents refrigerated at 208 °C
- Reagent Cartridge: 20 (with adaptors), 40 or 70 mL
- Dispensing system: Two reagent probes, 3-part reagent capability
- Reagent ID: Integrated Barcode reader
- Reagent volume: 5-300 uL
Reaction System:
- Reaction tray: 231 reusable high quality plastic cuvettes
- Reaction volume: 80 to 430 uL
- Temperature Reaction: 37 °C +/- 0.1 with inert bath oil
- Extended Reaction Time: 3, 4, 5, 10, 15 minutes
- Photometer reaction cycle: 21 or 31 minutes
- Homogenization 13.5 seconds
- Automatic correction: Rotation and repeated left-right movements by stirrers, serum blank, cell blank and measurement point change
Optical System:
- Light source: Halogen lamp with cooling system
- Principle wavelengths: Endpoint, rate reaction, 2 point rate, immunoassay 14 wavelengths (340-884 nm), mon and bichromatic
Physical specifications:
- Analyzer Size: 122 W x 85 D x 110.8 H cm or 48 in. W x 33 in. D x 43 in. H
- Analyzer weight: 450 kgs / 992 Lbs with AL 1208 Lbs
- Ambient Temperature: 18 – 30 °C
- Water Specifications: Conductivity < 1 us/cm
Specifications
Plasma, Serum or Urine | |
Processing Chemistry Tests | |
Automatic sample dilution 1/2 to 1/203 | |
Endpoint, rate reaction, 2 point rat, immunoassay | |
84 samples continuous loading | |
23 L/hr. | |
CE Marked, FDA Cleared | |
234 reusable high quality plastic cuvettes | |
Comprehensive test menu | |
24 | |
125 x 85 x 110.8 cm (WxDxH) / 48 x 33 x 43 in. (WxDxH) | |
85 cm | |
110.8 cm | |
122 cm | |
2.6 kVA | |
220 V | |
Yes | |
18°C to 30°C 40% to 70% with no condensation | |
Multi Cal, Protein Cal, Albumin U/CSF Cal, CRP Cl, RF Cal, Cystatin C Cal, Ferritin Cal, TPU Cal, HDL Cal, LDL Cal, hsCRP Cal, N multi, P multi, Protein L1,L2; Urine L1,L2;Albumin U/CSF L1,L2;CO2, Cystatin C L1,L2; Low CRPhs | |
Indirect | |
452 kg |
Na+, K+,Cl- and Reference electrode | |
12 minutes for one CMP panel 29 minutes for 20 CMP samples | |
1200 Tests/hr. | |
Spectrophotometry, Colorimetric and Turbidimetric | |
Yes | |
Floor Model | |
BOM List Analyzer, Computer, Monitor, Keyboard, Mouse, Power cords Printer Wash solutions Set of Electrodes ISE solutions and standards and USB key containing Daily guide, Operations manual | |
3 to 15 minutes, measurement cycle of 13.5 seconds | |
341/884 nm | |
Crates shipped on truck to hospital | |
33 in. | |
43 in. | |
48 in. | |
2.6 kVA, 220 V ± 10%, 50/60 Hz | |
50/60 Hz | |
Computer with keyboard | |
Yes | |
Windows Operating System | |
Operational | |
995 lb. |